(Ba,Sr)TiO3 - BASED ELECTROLYTE CONDUCTIVITY SENSORS: A COMPARATIVE STUDY

Authors

  • V.V. Buniatyan National Polytechnic University of Armenia Author
  • C.A. Huck 1FH Aachen | University of Applied Sciences, Institute of Nano- and Biotechnologies, Campus Jülich Author
  • A.S. Poghosyan 1FH Aachen | University of Applied Sciences, Institute of Nano- and Biotechnologies, Campus Jülich Author
  • M.J. Schöning 1FH Aachen | University of Applied Sciences, Institute of Nano- and Biotechnologies, Campus Jülich Author
  • T.N. Petrosyan National Polytechnic University of Armenia Author

Keywords:

contactless, electrolyte conductivity, perovskite-oxide film, interdigitated electrode.

Abstract

A contactless electrolyte conductivity sensor based on (Ba,Sr)TiO3 films and Pt interdigitated electrodes is fabricated and examined. Two types of sensor designs with buried(into insulating SiO2layer) and non-buried interdigitated electrodes were investigated and compared. The sensors were characterized by means of impedance-spectroscopy measurements indifferent commercially available electrolyte-conductivity standard solutions ranging from 0.084 to 50 mS/cm and over the frequency range from 1 Hz to 3 MHz. It is shown that the parameters of sensors with the Pt electrodes buried in SiO2 layer are favourable.

Downloads

Published

18.09.2026

Issue

Section

Articles

Most read articles by the same author(s)

Similar Articles

1-10 of 27